Link Biotech

A-1/264A Paschim Vihar, New Delhi - 110063

Vasopressin ELISA Kit

SKU Code LBEK-016U
Species Universal
Sensitivity 0.188pg/ml
Detection-range 0.313-20pg/ml
Size 96 T | 48 T
Price POR
Brand Link Biotech
SKU: LBEK-016U
Vasopressin ELISA Kit
UniProt ID P01185
Sample Volume Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples
Detection Method Competitive ELISA, Coated with Antigen
Detection Wavelength OD450
Reaction Duration 2 hours
Storage 2-8°C(Sealed), Don't cryopreserve.
Specificity Specifically binds with Vasopressin , no obvious cross reaction with other analogues.

Additional information

sensitivity

0.188pg/ml

detection-range

0.313-20pg/ml

Brand

Link Biotech

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(pg/ml) OD-1 OD-2 Average
0 2.135 2.244 2.179
0.312 1.671 1.757 1.706
0.625 1.444 1.518 1.474
1.25 1.198 1.259 1.223
2.5 1.021 1.073 1.042
5 0.683 0.717 0.697
10 0.47 0.497 0.483
20 0.257 0.27 0.263
EU3114 Standard Curve Image

Add a certain amount of Vasopressin into the sample. Calculate the recovery by comparing the measured value with the expected amount of Vasopressin in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 87-105 92
EDTA plasma(n=10) 88-102 94
Heparin plasma(n=10) 88-104 98

Dilute the sample with a certain amount of Vasopressin at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 94-101% 86-101% 91-99%
EDTA plasma(n=10) 82-94% 87-98% 82-91%
Heparin plasma(n=10) 82-98% 82-95% 90-96%

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/ml) 0.62 2.59 10.37 0.6 2.43 9.96
Standard deviation 0.04 0.14 0.33 0.03 0.11 0.54
CV(%) 6.42 5.23 3.21 4.88 4.68 5.43

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months 2-8°C for 12 months
Average(%) 80 95-100 85-98