Link Biotech

A-1/264A Paschim Vihar, New Delhi - 110063

Human ALD(Aldosterone) ELISA Kit

SKU Code LBEK-089H
Species Human
Sensitivity 1.875pg/ml
Detection-range 3.125-200pg/ml
Size 96 T | 48 T
Price POR
Brand Link Biotech
SKU: LBEK-089H
Human ALD(Aldosterone) ELISA Kit
UniProt ID CAS: 52-39-1
Sample Volume Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples
Detection Method Competitive ELISA, Coated with Antigen
Detection Wavelength OD450
Reaction Duration 2 hours
Storage 2-8°C(Sealed), Don't cryopreserve.
Specificity Specifically binds with ALD , no obvious cross reaction with other analogues.

Additional information

sensitivity

1.875pg/ml

detection-range

3.125-200pg/ml

Brand

Link Biotech

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(pg/ml) OD-1 OD-2 Average
0 2.111 2.173 2.142
3.125 1.872 1.926 1.899
6.25 1.333 1.371 1.352
12.5 0.91 0.936 0.923
25 0.604 0.622 0.613
50 0.413 0.425 0.419
100 0.269 0.277 0.273
200 0.163 0.167 0.165

EH4705 Standard Curve Image

Add a certain amount of ALD into the sample. Calculate the recovery by comparing the measured value with the expected amount of ALD in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 85-102 92
EDTA plasma(n=10) 88-103 92
Heparin plasma(n=10) 88-102 94

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/ml) 6.52 24.96 98.62 5.91 24.36 98.03
Standard deviation 0.35 1.17 4.91 0.34 1.25 4.57
CV(%) 5.33 4.69 4.98 5.74 5.12 4.66

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months 2-8°C for 12 months
Average(%) 80 95-100 85-98